This S-Class small steam sterilizer is equipped with the rectangular chamber and able to load the items to be sterilized in the chamber much more and conveniently. This is proper to sterilize effectively the general medical devices and supplies used frequently in the health care institutions.
1. Unique small steam generator
This sterilizer is equipped with the new type external steam generator which can vaporize water instantly, and the saturated vapor is injected into the chamber as a sterilizing media, therefore the chamber is less contaminated than the internal steam generating type and the usable chamber space is widened because there is no heater assembly in the chamber.
2. Use of clean steam every cycle
The used water after completing a sterilizing cycle is collected in the collecting container designed separately with the reservoir, therefore the saturated steam generated by a clean water always can be supplied to perform every sterilizing cycle.
3. Safe and durable unwelded chamber
By adopting a integral chamber unwelded, the durability is excellent and a trouble such as leaking of pressurized vapor and vacuum during progressing a cycle is not occurred accidently.
4. Ubiquitous and storing function of the cycle performance data (optional)
The cycle performance data executed for 10 years are stored in the control unit in order. User can download or backup those data in the USB memory and monitor the operation of sterilizer on the external PC by connecting the USB or LAN cable to the built-in communication ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilizers at once.
(These functions can be provided as an optional specification)
5. Various cycle programs of the optimum condition
5 kinds of the basic cycle program composed of the optimum sterilization parameters are installed in this sterilizer and the operator can select a proper cycle program among them conveniently for sterilizing, moreover user can change the set values of the basic cycle programs as user’s please if necessary for applying to a special sterilization condition or materials.
6. Safe and convenient self-diagnosis function
If an error has occurred during operating the sterilizer, it is detected and the error code is displayed on the LED display accompanying an audible alarm, and the cycle in progress is aborted automatically for safety by the built-in self-diagnosis program.
7. Convenient monitoring of a cycle processing
The operator is able to check a cycle progressing conveniently on the LED display because the state of sterilizer operation is displayed digitally in real time on it.
8. Printing of the cycle performance record (optional)
The external printer can be equipped to print out the cycle performance record for securing the traceability afterwards.
6(The printer is provided as an optional item (The exterior printer is provided as an optional item) |